Webster – How Blocked Gas Detectors Change the Apparent Concentration of Gas


Dr Ian Webster – Group Engineering Manager, Ampcontrol Pty Ltd

ABSTRACT

The operation of a diffusion type gas detectors used in fixed, machine mounted and handheld applications is reliant on the natural equalisation of dissimilar gas concentrations driven by partial pressures inside and outside the detector.

Typically, this equalisation is inhibited (to a greater or lesser degree) by protective filters and barriers surrounding the fragile sensing elements from the typically harsh ambient environments. The accumulation of dust and other foreign matter on the protective filters can further inhibit the diffusion of gas into a detector.

The usual calibration process for a gas detector – typically by a ‘bump’ or ‘challenge’ test – will often fail to detect when a detector is blocked, or partially blocked. This can lead to the ‘calibrated’ detector reading high or low, but with no way to determine if that is the case.

Retrospective examination of records and equipment from Pike River Mine lead to the conclusion that critical detectors were affected by filter blockages, resulting in methane detectors reading approximately one-half of the true concentration.

This presentation explores how a blocked detector can give an erroneous reading, and what steps can be taken to avoid replicating previous mistakes.

You may also like

Armitage - ISO 45003: Managing Psychosocial Hazards in the Workplace. What This Means for Your Business
28:59
Naomi Armitage, Director/Psychologist, Humanology Group Presentation at the 33rd Annual Queensland Mining Industry ...
Wilson - ISO Mining Standards to Protect Worker Health in Operator Enclosures
26:38
Liam Wilson, ISO Committee Member TC-82 Mining, WG 9, ISO Committee TC-82 – Mining, WG 9

Page 21 of 26